Brachyview: An in-body imaging system for real-time QA in HDR prostate brachytherapy
- Publication Type:
- Conference Proceeding
- Citation:
- IEEE Nuclear Science Symposium Conference Record, 2013
- Issue Date:
- 2013-01-01
Closed Access
Filename | Description | Size | |||
---|---|---|---|---|---|
06829797.pdf | Published version | 481.45 kB |
Copyright Clearance Process
- Recently Added
- In Progress
- Closed Access
This item is closed access and not available.
A transrectal real-time source tracking system, BrachyView, is developed by Centre for Medical Radiation Physics at University of Wollongong for treatment quality control during the high dose rate prostate brachytherapy procedure (HDR-PBT). The probe consists of a 15 × 60 mm2 pixellated silicon detector (four Timepix detectors) encased in a cylindrical seven pinhole tungsten collimator. The source position is determined by back-projecting the source projection images through the collimator pinholes. A prototype BrachyView probe consisting of a single Timepix detector (15 × 15 mm2) and a planar seven pinhole collimator is evaluated in this paper. The source position can be estimated with a maximum error of 0.2 mm on the detector plane, which translates into a maximum error of 1.42 mm within the prostate volume. This result is in good agreement with the results obtained in previously published simulation studies. © 2013 IEEE.