Localisation of the Lines of Response in a Continuous Cylindrical Shell PET Scanner
- Publication Type:
- Conference Proceeding
- Citation:
- Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS, 2019, pp. 4844 - 4850
- Issue Date:
- 2019-07-01
Open Access
Copyright Clearance Process
- Recently Added
- In Progress
- Open Access
This item is open access.
© 2019 IEEE. This work presents a technique for localising the endpoints of the lines of response in a PET scanner based on a continuous cylindrical shell scintillator. The technique is demonstrated by applying it to a simulation of a sensitivity-optimised continuous cylindrical shell PET system using two novel scintillator materials -a transparent ceramic garnet, GLuGAG:Ce, and a LuF3:Ce-polystyrene nanocomposite. Error distributions for the endpoints of the lines of response in the axial, tangential and radial dimension as well as overall endpoint spatial error are calculated for three source positions; the resultant distribution of error in the placement of the lines of response is also estimated.
Please use this identifier to cite or link to this item: