Developing computer controlled laser systems for double blind acupuncture trials
- Publication Type:
- Thesis
- Issue Date:
- 2012
Open Access
Copyright Clearance Process
- Recently Added
- In Progress
- Open Access
This item is open access.
Background: Double blind clinical testing is considered critical for advancing most
areas of medical research. Much of the acupuncture laser and Low Level Laser Therapy
(LLLT) research lacks double blinding and suffers from poor trial design. While some
research studies have shown a potential increased clinical effect in visible wavelength
laser trials visible wavelength lasers are generally overlooked in research, due to the
difficulties with double blinding.
Objectives: This study sought to design, construct and test a system that could
randomise, control, monitor and record the delivery of visible laser in real-time, and
allow for double blinding.
Method: The study utilised a novel approach to incorporate computer controlled lasers
within opaque hoods. An optical sensor feedback was incorporated to objectively
measure the delivery of laser in a double blinded randomised controlled trial (RCT)
situation.
The equipment was designed and tested for future researchers to operate with a
minimum of training.
The initial testing was designed around a proof of concept study using the equipment to
stimulate Pericardium 6 (PC 6) with a visible (635 nm) 10 mW laser while monitoring
changes in Heart Rate Variability (HRV). We only sought to trial the equipment under
laboratory conditions to see if a double blind condition could be achieved. It was
beyond the scope of this study to undertake a RCT of the effects of PC 6 on HRV.
Results: The equipment was successful in maintaining double blind conditions. Laser
was delivered and monitored in real-time, the operational parameters were recorded and
test participants reported no discomfort associated with the use of the equipment. The
computer successfully controlled the random delivery of laser. The test participants and
operators were at no time aware of the state of the laser. The operators reported no
difficulties in using the unit.
Conclusion: A laser research tool was successfully created and the study demonstrated
the system’s unique ability to overcome all the difficulties identified for visible laser
wavelength double blind research.
The system is currently designed with two laser hoods to allow for bilateral acupuncture
point testing. Future improvements could include additional laser hoods for multiple
acupuncture point treatments. Construction of alternate hoods would allow for testing
of different wavelengths including infrared and different laser power levels.
The system combines double blind testing with physiological monitoring, which allows
researchers to record the physiological effects that accompany laser therapies.
Please use this identifier to cite or link to this item: