© 2019 John Wiley & Sons, Ltd. The array of analytes that can be measured by MADLI MS has created an equally vast range of calibration mixtures. The inherent problem with this is that acquiring all of them at commercial rates can be prohibitively expensive. With this in mind, we have created a low-cost alternative to the most commonly used peptide calibrants. We were able to achieve an overall 78 ppm mass accuracy across a mass range of 900 to 2500 Da which was comparable to the mass accuracy achievable with commercial peptide mixes and hence has become a viable alternative.