Gas Cascade Amplification in Ultra-High-Resolution Environmental Scanning Electron Microcopy

Cambridge Univ Press
Publication Type:
Journal Article
Microscopy And Microanalysis, 2010, 16 (6), pp. 805 - 809
Issue Date:
Full metadata record
Files in This Item:
Filename Description SizeFormat
2010004242OK.pdf192.93 kBAdobe PDF
We describe a feedback mechanism in the gas cascade amplification process used in magnetic immersion lens environmental scanning electron microcopy (ESEM). Feedback dominates gas gain under the conditions typically used for ultra-high-resolution ESEM and gives rise to novel dependencies of the imaging signal and noise on microscope operating parameters. It is ascribed tentatively to the generation of free electrons upon de-excitation of metastable species in the gas cascade. The results have implications for optimization of ESEM systems for applications such as critical dimension metrology and real-time imaging of nanostructure growth by gas mediated electron beam induced deposition.