Improving acquisition times of elemental bio-imaging for quadrupole-based LA-ICP-MS

Royal Society of Chemistry
Publication Type:
Journal Article
Journal of Analytical Atomic Spectrometry, 2012, 27 (1), pp. 159 - 164
Issue Date:
Full metadata record
Files in This Item:
Filename Description SizeFormat
2011003874OK.pdf559.82 kBAdobe PDF
Elemental bio-imaging experiments by quadrupole-based LA-ICP-MS normally employ scan speeds where the distance traversed in one second is equal to or less than the diameter of the laser beam. Consequently, data for a higher-resolution (pixel size = 15 µm2) image of a 5 mm2 tissue section can take upwards of 30 h to acquire. Appropriate laser scan speeds may be calculated by consideration of the relationship between laser scan speed, laser spot diameter and the total scan cycle of the quadrupole mass analyser. This paper presents a simple method to calculate the laser scan speeds capable of reducing the acquisition time by up to a factor of 5 whilst maintaining dimensional integrity of the image.
Please use this identifier to cite or link to this item: